Table 4-3. Organizational Preventive Maintenance Checks and Services-continued
Up
TM-11-6625-444-14-2 Voltmeter Digital AN/GSM-64C (NSN 6625-01-124-0834) (EIC: KNK) Manual
Next
Functional Testing

TM 11-6625-444-14-2
Section IV. TROUBLESHOOTING
it off and check the following items:
4-6. General
This section contains troubleshooting instruc-
tions for the voltmeter. Any malfunction that is
a. Wrong control settings or improper input
beyond the scope of operator or organizational
connections (para 3-3, 3-6, 3-7, 3-8, and 3-9).
maintenance shall be referred to direct and gen-
eral support maintenance in Chapter 6.
b. Damaged or incorrectly installed power
cord.
4-7. Procedure
c. Defective power fuse (figure 2-2).
If the voltmeter fails to operate correctly, turn
Section V. MAINTENANCE OF VOLTMETER
(4) Grasp the right end of the fuse and pull
4-8. General
it out from the left spring clip.
This section contains maintenance instructions
applicable to operator and organizational activi-
To install a replacement power fuse, proceed
ties. Maintenance of the voltmeter by these ac-
as follows:
tivities is limited to the following items:
(5) Rotate the FUSE PULL handle back to
a. Fuse and knob replacement.
its normal (fully right) position.
b. Cleaning (para 4-10).
(6) insert the new fuse fully into the spring
clips.
c. Functional testing (para 4-11).
4 - 9 . Removai/Replacement
CAUTION
a. Fuse Replacement. The power fuse F1 is
To avoid damage to the voltmeter,
located within the ac power connector assembly
replace defective fuse with one
on the rear panel (figure 3-2).
having identical rating.
To remove the power fuse, proceed as follows:
(7) Slide the cover back to its normal (ex-
(1) Disconnect the power cable from the
treme right) position.
fused receptacle.
(8) Reconnect the power cable to the fused
(2) Slide the clear plastic cover to the ex-
receptacle.
treme left position (figure 2-2).
b. Knob Replacement. The SAMPLE RATE/-
(3) Grasp the FUSE PULL handle and swing
HOLD and RATIO FULL RANGE knobs are re-
it fully to the left. This will disconnect the right
placed as follows:
end of the fuse from its spring clip retainer.

Privacy Statement - Press Release - Copyright Information. - Contact Us