Page Title: Table 1-3. Ac Voltage Accuracy
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
Table 1-2. Ac Voltage Ranges*
Up
TM-11-6625-444-14-2 Voltmeter Digital AN/GSM-64C (NSN 6625-01-124-0834) (EIC: KNK) Manual
Next
Figrue 1-2. Outline Drawing, Portable Configuration

TM 11-6625-444-14-2
Table 1-3. Ac Voltage Accuracy
ACCURACY **
FREQUENCY
± (0.1% of reading + 400 counts)
dc (coupled)
± (0.2% of reading + 600 counts)
0.5 Hz to 1 kHz
± (0.2% of reading + 400 counts)
1 kHz to 30 kHz
± (0.8% of reading + 400 counts)
30 kHz to 100 kHz
± (1.5% of reading + 600 counts)
100 kHz to 300 kHz
± (3.0% of reading + 4000 counts)
300 kHz to 1 MHz
** Applies from 20,000 counts to 199,999 counts over a 90 day period at + 23°C ±2°C.
Relative humidity less than 90%.
shunted by
(a) Input impedance: 1 M
(4) Shock and Vibration: Meets require-
ments of MIL-STD-810; Procedure IX, Method
less than 50 pF (less than 100 pF rear input)
514 and Procedure V, Method 516
(b) Temperature coefficient: ±(0.01%
(5) Ac voltage:  100/120/220/240 V ±10
rdg + 6 LSD + 3 uV)/°C Up to 10 kHz. ±(0.02%
% single phase, sinusoidal 50 to 400 Hz
rdg + 10 LSD + 3 uV)/°C for 10 kHz to 1 MHz.
(6) Power: 50 watts
(c) Low pass filter: -3 dB at 10 kHz
(7) Display:  5½ digit, ± polarity sign, 7
300
(to 40 Hz)
(d) Response time:
s e g m e n t LED, 0.43 inch (11 mm) character
30 sec (to 0.5 Hz) slow.
normal.
height, characters bright red
90 dB
(e) Common mode rejection:
at 60 Hz
c. Mechanical
(f) Crest factor:  Full range (199,999
3.50 inches (88.9 mm)
(1) Height:
counts) 5 to 1. Half range (100,000 counts) 10
to 1.
16.75 inches (425.5 mm)
(2) Width:
(3) D e p t h : 17.12 inches (434.8 mm) be-
b. General
hind panel 17.625 inches (447.7 mm) overall.
(1) Temperature:
0 to 50°C operating.
(4) Weight: 18 lbs.
-55°C to +75 C storage.
(2) Humidity: Meets requirements of MIL-
STD-810; Procedure II, Method 507
1-8. Hems Comprising an Operable Equipment
(3) Altitude:  Meets requirements of tvllL-
Digital Voltmeter AN/GSM-64C, power cable
STD-810; Procedure 1, Method 500
and test leads comprise an operable equipment.