Section II. DESCRIPTION AND DATA
Up
TM-11-6625-444-14-2 Voltmeter Digital AN/GSM-64C (NSN 6625-01-124-0834) (EIC: KNK) Manual
Next
Table 1-3. Ac Voltage Accuracy

TM 11-6625-444-14-2
(a) Input impedance: >10,000 M
(c) Temperature coefficient: 20.0007%
of Rdg./°C
on 2V Range.
on 20V, 200 V and 1000 V
Ranges.
(d) External reference voltage: ±1 V
dc to ±10 Vdc to ±100 Vdc auto ranging on
(b) Input bias current:
100 picoam-
20V, 200V and 1000V Ranging; ±1V to ±2V on
peres at 20°C
2V Range.
(c) Normal mode rejection: >60 dB at
60 Hz (Filter In)
(e) Polarity of EXT REF voltage: Auto
Polarity on all ranges
(d) Common mode rejection (with driv-
Real Time 4
(f) Operation:
True -
120 dB at 60 Hz,
140 dB at dc
en GUARD):
wire ratio
unbalance in either lead.
with
(e) Polarity Indication: Automatic
(g) Common mode voltage (EXT REF
LO and INPUT LO): ±12 volts peak with EXT
REF  ±1V  to  ±11.5  V.  ±30  volts  peak  with
Temperature coefficient: 20.0007%
EXT REF ±11.5 V to ±100 V.
of Rdg./°C
resistance: > 1
2 seconds (Fil-
(g) Response time:
(h) Common mode
ter Out)
Meg ohm
2.2 M
on all
(i) Input indication:
ranges
(2) Dc Voltage Ratio
Ratio range: ±0.001% to ±100.000%
2 seconds
(k) Response time:
to ±10000.0%
(3) Ac Voltage (True rms)
(b) Ratio accuracy: ± (0.005% of Rdg.
+1 LSD)
Table 1-2. Ac Voltage Ranges*
RESOLUTlON
MAX. RDG.
RANGE
10 UV
1.99999 v
2V
100 UV
19.9999 V
20 V
1 mV
200 V
199.999 V
10 mV
1000.00 V
1000 V
*Maximum signal input (dc + ac rms) between HI and LO terminals iS limited to
1 08 V H z .