TB 9-6625-396-50
a. Performance Check
(1) Set G switches to 0000.
(3) Set BRIDGE READ switch to the middle unmarked position between SERIES
(5) Set G switches to 9999.
(7) Subtract value recorded in (4) above from value recorded in (6) above.
Difference will be between 0.98770I and 1.009899 F.
(8) Repeat technique of (5) through (7) above for G switch settings listed in table 5.
Capacitance value will be within limits specified.