SECTION II EQUIPMENT REQUIREMENTS
Up
TB-55-4920-346-34 Calibration Procedure for Environment Control Test Set NSN 4920-00-111-5372 GRUMMAN MODEL A/E24T-102 Manual
Next
Equipment Setup

TB 55-4920-346-34
Table 3. Accessories Required
Common Name
Description
Item
(Official Nomenclature)
(Part Number)
B1
Cable Assembly
36-in. RG-5810 BNC to double banana plugs
B2
Test lead
1
24-in. No. 18 lead single banana plugs
terminations (red)
B3
Test lead
1
24-in. No. 18 lead single banana plugs
terminations (black)
B4
Adapter
1
Banana Jack to pin tip (1432 = 0)
B5
Adapter
2
Banana Jack to alligator clip (2240-0)
B6
Adapter
Test Prod #6 Banana Jack to miniature hook
(4724-B)
B7
Adapter connector
2
Banana plug to cannon plug pin (3560 black)
B8
Adapter connector
Banana plug to cannon jack 3561-0 (red)
B9
Adapter box
SK-D-4850-3
1 Three required
2 Four required
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration process. Personnel should
become familiar with the entire bulletin before beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by common name and item identification
number as listed in tables 2 and 3. For the identification of equipment referenced by item numbers prefixed with A, refer
to table 2, and for prefix B, refer to table 3.
HIGH VOLTAGE is used or exposed during the performance of this calibration.
DEATH ON CONTACT may result if personnel fail to observe safety precautions.
4