HARMONIC DISTORTION
A magnetic tape recorder's harmonic-distortion specification is very important. It
usually determines where the record level of a recorder's electronics should be set. The
record level is also used to determine the signal-to-noise ratio and frequency-response
specifications. A typical harmonic-distortion specification might read "1% third
harmonic of a 100-kHz signal at 60 ips." This means that the magnetic tape recorder
has 1% third-harmonic distortion of a 100-kHz signal at 60 ips.
HARMONIC-DISTORTION DEFINITION
Harmonic distortion is the production of harmonic frequencies by an electronic
system when a signal is applied at the input.
When an input signal goes through nonlinear electronic circuitry, the output signal
will include some harmonic distortion (or unwanted frequencies).
If you analyzed this distortion, you'd see that a pattern exists. A pattern, whereby
the frequency of each unwanted frequency is a multiple (X1, X2, X3, etc.) of the center
frequency of the input signal.
There are two types of harmonic distortion: even-order and odd-order.
If the frequencies of the distortion are 2, 4, 6, etc., times the center frequency,
it's even-order harmonics. If the frequencies of the distortion are 3, 5, 7, etc., times
the center frequency, it's odd-order harmonics.
Odd-order harmonics are normally caused by the magnetic tape itself.
Even-order harmonics are normally caused by (1) permanently magnetized magnetic heads,
(2) faulty circuits, or (3) asymmetrical or unbalanced bias signals.
As you might guess, even-order harmonics can be reduced by doing the right maintenance
and periodic performance tests.
The primary harmonic distortion in magnetic tape recorder systems is third-order
harmonics. If the level of third-order harmonics in a recorder increases, the level of
distortion will also increase (figures 6-4A and B show this relationship). Two things that
determine the level of third-order harmonics in a recorder are (1) the signal bias level,
and (2) the record level. Figure 6-4A shows how third-order harmonic distortion decreases
as the signal bias level increases. Figure 6-4B shows how the third harmonic increases
gradually at first and then abruptly as the record level increases. That's why the third
harmonic is used to determine the normal record level.
Figure 6-4 A & B. - Effect of signal bias level and record level on
harmonic-distortion level.
HARMONIC-DISTORTION MEASUREMENT
Figure 6-5 shows a typical test equipment setup for measuring harmonic distortion. With
this setup, the test signal from the signal generator is recorded and reproduced by the
magnetic tape recorder at a normal record level. The amount of harmonic distortion is
measured at the recorder's output on the wave analyzer.
Figure 6-5. - Test equipment setup for measuring harmonic distortion.
The technical manual for the magnetic recorder you're testing will tell you how to set
up the test equipment. It'll tell you to set up the wave analyzer to measure a specific
frequency. This frequency will be one of the multiples (X1, X2, X3, etc.) of the frequency
the signal generator is outputting.
For example, let's say the technical manual told you to set up the signal generator to
input a 10-kHz test signal into the magnetic tape recorder. Since you want to measure
third-order harmonics, the technical manual will tell you to set the wave analyzer to
measure the amount of harmonic distortion at 30 kHz.
PHASE RESPONSE
It used to be thought that the only important specifications of magnetic tape recorders
were signal-to-noise ratio and Frequency response. But now, with the need to record and
reproduce more complex waveforms, such as telemetry and computer data, the phase-response
specification becomes as important as Frequency response.
PHASE-RESPONSE DEFINITION
Phase response is the expression of the variation of the phase shift with respect to
frequency. A good magnetic tape recorder will have linearly increasing phase response
as frequency increases.
In simpler terms, good phase response shows that a magnetic recorder can reproduce a
complex waveform (such as a square wave which has an infinite number of sine waves)
without distorting it. Figure 6-6 shows both good and bad phase response.
Figure 6-6. - Pictures showing the effect of good and bad phase response on square-wave
reproduction.
PHASE-RESPONSE MEASUREMENT
You cannot directly measure phase response. The best way to check the phase
response of a magnetic tape recorder is to record and reproduce a square wave and watch
the output on an oscilloscope.
If the output signal is symmetrical, like in figure 6-7,the recorder has good phase
response.
Figure 6-7. - An example of good linear phase response.
Q.7 A recorder's harmonic-distortion specification reads 2% third harmonic of a 100-kHz
signal at 60 ips. What does this mean? 
Q.8 What are three possible causes of third-order harmonics? 
Q.9 What number harmonic is the primary harmonic distortion in magnetic tape recorders? 
Q.10 When measuring harmonic distortion, you set the signal generator to input a 15-kHz
test signal. To what frequency should you set the wave analyzer? 
Q.11 How should a tape recorder with good phase response reproduce a complex
waveform, such as a square wave?
Q.12 How could you check the phase response of a tape recorder?