TB 9-6625-2233-35
(a) CALIBRATE (allow enough time to complete and verify AUTOCALIBRATED
is displayed).
Figure 4. A7
phase lock loop circuit card adjustment location.
(b)
RESET.
(c)
AMPLITUDE.
(d)
1. 1 1.
(e)
ENTER.
(f)
(g)
50.
(h)
ENTER.
(3) Multimeter will indicate between minimum and maximum values listed in table 6.
(4) Repeat technique of (2) (c) through (h) and (3) above substituting settings and
Table 6. Sine Wave Amplitude Accuracy 50 Hz to
99.99 kHz
Test instrument
Output
Min
Max
(V p-p)
50
Hz
1.11
0.381
0.404
50
Hz
.099
.0339
.0361
50
Hz
.2
.06858
.07282
50
Hz
.999
.345
.361
50
Hz
2
.689
.725
50
Hz
15
5.14
5.46
99.9 kHz
1.11
0.381
0.404
99.9 kHz
.099
.0339
.0361
99.9 kHz
.2
.06858
.07282
99.9 kHz
.999
.345
.361
99.9 kHz
2
.689
.725
99.9 kHz
15
5.14
5.46
16