TB 9-6625-1314-35
(8) Adjust electrical test set to obtain an indication of 199.9 on TI. Electrical test
set will indicate between 199.10 and 200.70 mA.
(9) Press TI 2000 range pushbutton.
(10) Adjust electrical test set to obtain an indication of 1999 on TI. Electrical test set
will indicate between 19830 and 20150 mA.
b. Adjustments. No adjustments can be made.
29. Ac Volts (Level A and AN/GSM-259)
a. Performance Check
(1) Connect ac voltage calibrator (A2) to TI VOLTS and COM terminals using cable
(B2).
(2) Press TI ac function and 2 range pushbuttons.
(3) Adjust ac voltage calibrator frequency to 100 Hz and voltage to obtain an
indication of 1.999 on TI. If ac voltage calibration does not indicate between 1.9790 and
2.0190 V ac, perform b(1) and (2) below.
(4) Repeat technique of (2) and (3) above using settings and indications listed in
table 16. Ac
voltage calibrator will
indicate within
limits specified.
Table 16. Ac Volts (Level A and AN/GSM-259)
Test instrument
RANGE
pushbutton
Indication
Min
Max
2
1.999
50 Hz
1.9790
2.0190
2
1.999
1 kHz
1.9790
2.0190
2
1.999
10 kHz
1.9790
2.0190
2
1.999
20 kHz
1.9510
2.0470
2
1.999
100 kHz
1.8870
2.1110
.2
199.9
100 Hz
0.1979
0.20191
.2
199.9
50 Hz
0.1979
0.2019
.2
199.9
1 kHz
0.1979
0.2019
.2
199.9
10 kHz
0.1979
0.2019
.2
199.9
20 kHz
0.1951
0.2047
.2
199.9
100 kHz
0.1887
0.2111
20
19.99
2 kHz
19.790
20.1902
20
19.99
50 Hz
19.790
20.190
20
19.99
100 Hz
19.790
20.190
20
19.99
1 kHz
19.790
20.190
20
19.99
10 kHz
19.790
20.190
200
199.9
50 Hz
197.90
201.90
200
199.9
100 Hz
197.90
201.90
33