![]() ![]() TB 55-4920-346-34
Figure 7. Simulator Equipment Setup
Table 6. Temp Control and Sensor Simulators Check
Test Instrument Temp and
Resistance Voltmeter Indication (Ohms)
Sensor Simulator Settings
Min
Max
00100.0
99.50
100.50
00250.0
248.75
251.25
00500.0
447.50
502.50
01000.0
995.00
1005.00
05000.0
4975.00
5025.00
10000.0
9950.00
10050.00
50000.0
49750.00
50250.00
90000.0
89550.00
90450.00
13. Final Procedure.
a. Deenergize and disconnect all equipment and install TI into case and reinstall the protective cover.
PIN: 0093291001
*U.S. Government Printing Office; 1989-654-031/55043