Click here to make tpub.com your Home Page

Page Title: Table 5. D.C. Milliamp
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
Table 4. AC Milliamp
Up
TB-55-4920-346-34 Calibration Procedure for Environment Control Test Set NSN 4920-00-111-5372 GRUMMAN MODEL A/E24T-102 Manual
Next
Final Procedure.

TB 55-4920-346-34
Table 5. D.C. Milliamp
Meter Calibrator Indication (MA)
Test Instrument Indication (MA)
Min
Max
200
175
225
300
275
325
400
375
425
500
475
525
12. Simulators (Temp and Sensor).
a. Performance Check. (Figure 6 and Figure 7)
(1) Connect equipment as shown in figure 6. Check test leads resistance and record.
(2) Set the TEMPERATURE CONTROL SIMULATOR to 00100.0 on the test instrument.
(3) The voltmeter (A2) shall indicate within the limits specified in table 6, after substracting the lead resistance
recorded in (1) above.
(4) Repeat the above check for each of the test instrument TEMPERATURE CONTROL settings listed in table 6.
(5) Disconnect test leads from TP-31 and 32 and connect to TP-39 and 40.
(6) Set the SENSOR SIMULATOR to 00100.0 on the test instrument.
(7) Repeat steps (3) through 4 above, using SENSOR CONTROL settings in table 6.
b. Adjustments. No adjustments can be made.
Figure 6. DC Milliamp Equipment Setup
Change 1
11