4-26. LEAK TEST PROCEDURE
Up
TM-5-6635-386-12P Density and Moisture Tester Manual
Next
4-28. RECORDS

TM 5-6635-386-12&P c.  Carefully  place  the  tester  on  its  end  or  side. (See Fig. 4-2). d. Moisten the cotton swab in the detergent/water solution. e. Swab the ring which is located on the bottom of the tester. Do not swab the source rod or the shutter. (See Fig. 4-2). f.  Place  the  cotton  swab  on  the  sheet  of  plastic wrap. g.  Return  the  tester  to  its  upright  position.  (See Fig. 4-1). Figure 4-2. Swabbing the Ring h.  Loosen  the  four  screws  which  are  located  in the corners of the tester’s display panel. Lift the panel out of the tester and support it on the case.  (See Fig. 4-3). i. With the same swab that was used in e above, swab the red spot which is located within the tester and next  to  the  internal  radiation  detector.  The  spot  to  be swabbed  may  not  be  painted  red  in  some  testers.  (See Fig. 4-3). j. Wrap just the cotton swab and an inch or two of   the   stick   with   plastic   wrap.   Seal   the   plastic   wrap around the stick with cellophane tape. Figure 4-3. Swabbing the Spot k.  Reinstall  the  display  panel  in  the  tester  and tighten the four screws. I.  The  swab  shall  be  tested  for  radiation  in  an area where background radiation averages 0.04 mrem/hr or less. (1)    Adjust    the    radiac    meter    AN-PDR/27    or equivalent to measure 0 to 0.5 mrem/hr. (2)   Open   the   cover   end   of   the   radiac   meter’s probe. (3) Do not touch the swab to the meter’s probe. Place the swab about 0.25 inch in front of the probe and note the meter reading. WARNING A  tester  is  contaminated  when  a  reading  of  0  I mrem/hr or more (or in the case of Colorado 0.2 mrem/hr  or  more)  is  obtained  from  the  swab. Follow  the  procedures  provided  in  Para    4-31. The swab is also contaminated. The  tester  shall  immediately  be  withdrawn  from use when the wipe test indicates a reading of 0.1 mrem/hr.    (or    in    the    case    of    Colorado    0.2 mremlhr)   or   when   the   laboratory   test   results reveal the presence of 0.005 or more microcuries. m.  Cut off the applicator stick at the tape seal. n.    If    a    swab    reading    greater    than    0.4    mrem/hr 4-9

Privacy Statement - Press Release - Copyright Information. - Contact Us - Support Integrated Publishing