Page  Title: 4-26. LEAK TEST PROCEDURE
Back | Up | Next

Click here for a printable version

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Combat
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
USMC
   
Products
  Educational CD-ROM's
Printed Manuals
Downloadable Books
   

 

Back
SURVEYS AND TESTS
Up
TM-5-6635-386-12P Density and Moisture Tester Manual
Next
Figure 4-2. Swabbing the Ring

TM 5-6635-386-12&P LRPO, and shall be recorded in microcuries. d. Wipe tests made by users in the U.S., Puerto Rico, Panama and Korea shall be sent to: Chief, us Army Ionizing  Radiation  Dosimetry  Center,  AMXTM-S-LR-DN, Lexington, KY 40511-5102. e. Wipe tests made by USAREUR units shall be sent   to:   Radiac   ACL,   Pirmasens,   524th   Maintenance (TMDE) Co., APO NY 09138-4628. f. Overseas users in other than USAREUR units will  be  notified  of  Army  laboratories  to  which  wipe  tests shall be sent. g.   Units   that   receive   leak   test   kits   every   six months are not required to send test results to TACOM. The  test  results  are  forwarded  to  TACOM  by  the  U.S. Army Ionizing Radiation Dosimetry Center. h. Users who do not receive leak test kits every six    months,    shall    send    test    results    to    Commander, TACOM,  ATTN:  AMSTA-CZ.  Test  results  shall  include the  LRPO’s  name,  and  a  statement  that  the  tester  is being used only by trained personnel. (1) The LR PO shall ensure that test results are recorded   and   submitted   on   DA   Form   3252-R   (Punch Card   Transmission   Worksheet   Radioisotope   Inventory and  Leak  Test  Report),  as  required  by  AR  385-11  and Para 4-9e. (2)  When  filling  out  the  DA  Form  3252-R,  use the  last  five  digits  of  the  tester  serial  number  for  blocks 18-22. i. The LRPO shall immediately withdraw a tester from    use    when    results    of    leak    testing    reveal    the presence   of   0.005   or   more   microcuries.   Immediately thereafter,  the  LRPO  shall  notify  the  installation’s  RPO, the RCO, and the Commander, TACOM. The telephone shall be used to notify TACOM. A request for instructions from  TACOM  shall  also  be  made  during  the  telephone call.  TACOM (AMSTA-CZ) telephone numbers are: (1)   During   office   hours:   DSN    7866121/6194, Commercial (313) 574-6121/6194. (2) After office hours: DSN 786-5511, Commercial (313) 574-5511. 4-26. LEAK TEST PROCEDURE a. The tester contains two sealed sources. Both of  the  sources  must  be  tested  for  radiation  leakage  to comply with AR 385-11. See Para 4-25. b. The following items are required to performing the leak test: (1) Transparent plastic wrapping material. (2) Pressure sensitive transparent tape. (3) Cotton swab with wooden stem. (4) Detergent and water solution. (5) Aluminum or other metallic foil. (6) Radiac meter AN-PDR/27 or equivalent. 4-27. LEAK TEST a.  Place  a  sheet  of  clean   plastic   wrap   on   an uncontaminated surface near the tester. b. Make sure the testers handle is locked in the "SAFE" position, and that the shutter is closed. (See Fig. 4-1). Figure 4-1. Model MC-1 Nuclear Method Density and Soil Tester 4-8

Privacy Statement - Press Release - Copyright Information. - Contact Us - Support Integrated Publishing