![]() ![]() Section V
TM 11-6625-2781-14&P-2
ADJUSTMENT PROCEDURES (cont' )
d
5-25. 150 MHz Oscillator Adjustment
REFERENCE: Service Sheet 9.
DESCRIPTION: The 150 MHz oscillator is adjusted with a nonmetallic screwdriver to 150 MHz ±10 kHz. the specified
operating frequency, output amplitude of the oscillator should be 150 to 230 millivolts rm During the adjustment caution
must be used to avoid damaging the tuning slug.
Figure 5-6. 150 MHz Oscillator Adjustment Test Setup
EQUIPMENT:
Vector Voltmeter ................................ ................................ ................................ ................................ ....... HP 8405A
Frequency Counter ................................ ................................ ................................ ......................... HP 5245L/5252A
Extender Cable Assembly ................................ ................................ ................................ ............... HP 11592-60015
Interconnecting Cable Assembly ................................ ................................ ................................ ..... HP 11592-60016
Cable Assembly ................................ ................................ ................................ .............................. HP 11592-60001
BNC Cable Assembly................................ ................................ ................................ .............................. HP 10503A
50-Ohm Tee ................................ ................................ ................................ ................................ ........... HP 11536A
Adapter, Male Type N to BNC Female ................................ ................................ ................................ .... UG-201A/U
10 dB Attenuator Pad................................ ................................ ................................ ................................ HP 8491A
Nonmetallic Tuning Tool
Type N Female to BNC Female Adapter ................................ ................................ ................................ ..FXR 21850
1. Connect test setup in Figure 5-6. Make the following control settings.
ANALYZER:
RANGE MHz ................................ ................................ ................................ ................................ ................... 0-110
FREQUENCY ................................ ................................ ................................ ................................ ................. 0 MHz
5-12
|
|
Privacy Statement - Press Release - Copyright Information. - Contact Us |