Figure 3-1. Spectrum Analyzer Controls, Indicators and Connectors
Up
TM-11-6625-2781-14-P-2
Next
Figure 3-1. Spectrum Analyzer Controls, Indicators and Connectors - continued

TM 11-6625-2781-14&P-2
Section III
intensity baseline).
Blanking function also
prevents
blooming
with
a
variable-
per division absolute voltage amplitude (la) is product
persistence/storage display section.
(blue numerals) of LINEAR SENSITIVITY controls.
21. Adjusts vertical position and gain of trace.
33. Provides 1- and 10-volt, peak-to-peak, 60 Hz
squarewave outputs.
22. Adjusts horizontal position and gain of trace.
CAUTION
These calibrated outputs must never
23. Indicates  1-dB  increments  for  logarithmic
be used with the spectrum analyzer.
amplification; indicates multiplication factors up
(These outputs are for use only with
to unity for linear amplification.
the 1400-series oscilloscope plug-
ins).
24. Selects scan trigger mode.
34. Makes base line parallel with the horizontal
25. Selects logarithmic or linear display mode.
graticule line.
26. Assuming that dB graduation (black numerals)
35. Adjusts brightness of CRT display.
matches position of lighted index lamp, LOG
REF graticule line indicates power level when
CAUTION
LOG/LINEAR (25) is set to LOG. With LOG/
Excessive brightness for a static or
LINEAR set to LINEAR, indicates per division
very slow moving trace may burn the
multiplier for calibrated voltage amplitude for
phosphor and permanently damage
whatever voltage graduation (blue numerals)
the CRT. This caution is applicable
matches position of lighted index lamp.
to  both  the  fixed  and  variable
persistence/storage CRT; however,
27. Selects scan ramp mode.  Ramp is internally
the latter is especially vulnerable to
generated for SINGLE/INT positions but it must
operational errors of this type.
be externally supplied for EXT position. (Refer
to Item 19).
36. Used with FOCUS control (37) to obtain
smallest spot with maximum roundness.
28. Press to initiate scan with SCAN MODE switch
set to SINGLE; press during scan to stop and
reset scan.
37. Focuses CRT beam.
29. Controls SCAN TIME.
38. Lights  when  line  voltage
is
applied
and
instrument is turned on.
30. May select 100 Hz, 10 kHz or OFF position of
low-pass filter for detected video.
39. Switches line voltage to instrument.
31. Lights for duration of each scan for SINGLE and
40. When used with 1400-series oscilloscope plug-
INT scan modes.
ins, intensifies and returns beam to CRT,
regardless of deflection potentials. When used
with 8550 series plug-ins, the beam finder has
32. Blanks  lower  part  of  trace  to  prevent
no function.
overexposure of photographs (due to high
3-4