TB-9-6625-2252-35 Electronic Voltmeter ME-202C/U Manual
Next
Equipment Setup

TB 9-6625-2252-35
5. Accessories Required.  The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Fluke, Model 5720A (5700A/EP)
CALIBRATOR
Dc volts:
(p/o MIS-35947) w/amplifier, Fluke,
Range: 1 to 100 V dc
Model 5725A/AR (5725A/AR)
Accuracy: ±.0125%
Ac volts:
Range: 1 to 1000 V ac
Frequency: 200 Hz to 20 kHz
Accuracy:  ±.05%
MULTIMETER
Range: -0.1 to 2.6 V dc
Fluke, Model 8840A/AF05
(AN/GSM-64D)
Accuracy: ±1.0%
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in TM 11-6625-2724-12 for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.