![]() ![]() TB 9-6625-2182-35
EQUIPMENT REQUIREMENTS
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-287. Alternate items may be used by the calibrating activity. The
items selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
standard and TI. Where the four-to-one ratio cannot be met, the actual accuracy of the
equipment is shown in parenthesis.
accessories, issued as indicated in paragraph 4 above, and are not listed in this calibration
procedure. The following peculiar accessories are also required for the calibration:
Semiconductor Device (coaxial crystal detector), Hewlett-Packard, Model 423AOPT03.
Manufacturer and model
Common name
Minimum use specifications
(part number)
AUDIO ANALYZER
Frequency measurement:
Boonton, Model 1121 (1121)
Range: 20 Hz to 100 kHz
Accuracy: +0.5%
Distortion capability: <.02%
AUTOTRANSFORMER
Range: 105 to 125 V ac
Ridge, Model 9020A (9020A)
Fluke, Model PM6681/656
FREQUENCY
Range: 20 Hz to 1900 MHz
(PM6681/656)
COUNTER
Accuracy: +2.5 ppm or .00025%
MEASURING
Power measurement: (+15 dB to -110 dB) +.5 dB
Hewlett-Packard, Model 8902A
RECEIVER
Flatness measurement:
w/sensor module, Hewlett-Packard,
(100 kHz to 450 MHz) +.375 dB
Model 11722A (11722A)
(550 MHz to 950 MHz) +.5 dB
(1500 MHz to 2000 MHz) +.625 dB
MULTIMETER
Range: 50 to -15 V dc
Fluke, Model 8840A/AF05
Accuracy: +.25%
(AN/GSM-64D)
OSCILLOSCOPE
Range: 50 Hz
(OS-303/G)
Accuracy: <125 ns risetime
PULSE GENERATOR
Amplitude: 5 V
LeCroy, Model 9210 (9210)
Period: 10 ms to 20 µs
w/plug-ins, LeCroy, Model 9211
Width: 5 ms to 6 µs
(9211)
SPECTRUM ANALYZER
Range: 450 kHz to 2 GHz (13 to -90 dB)
(AN/USM-677)
Accuracy: +1.0 dB/10 dB step, 1.0 dB maximum
|
|