![]() ![]() TB 9-6625-2182-35
IDENTIFICATION AND DESCRIPTION
This bulletin provides instructions for the
calibration of Signal Generator, SG-1207/U (Hewlett-Packard, Model 8642M). TM 11-6625-
3165-14 was used as the prime data source in compiling these instructions. The equipment
being calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 6
hours, using the dc and low frequency and microwave technique.
personnel at all levels are prescribed by TB 750-25.
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Frequency
Range: 100 kHz to 2000 MHz
Accuracy: +10 PPM
Time stability: +0.05 PPM/hour1
Line stability: +0.05 PPM, 10% line change2
RF output
Range: >15 to -140 dBm3
Flatness:
+1.5 dB (100 kHz to 500 MHz)
+2 dB (500 to 1000 MHz)
+2.5 dB (1 to 2 GHz)
Attenuator accuracy: +2.0 dB
Spectral purity
Harmonics range
Accuracy
100 kHz to 1057.5 MHz
<+13 dBm, <-25 dBc
1057.5 to 2000 MHz
<+7 dBm, <-25 dBc
Sub harmonic range
Accuracy
100 kHz to 1057.5 MHz
<-100 dBc
1057.5 to 2000 MHz
<-45 dBc
Spurious signal range
Accuracy
100 kHz to 132.1875 MHz
<-70 dBc
132.1875 to 1057.5 MHz
<-90 dBc
See footnote at end of table.
|
|