Click here to make tpub.com your Home Page

Page Title: ICES and IR Measurement Tests
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
Transconductance Test (FET In-Circuit)
Up
TM-11-6625-539-14-4 Test Set Transistor TS-1836 D/U Manual
Next
Ico lR Measurement Tests

TM 11-6625-539-14-4
Control settings
Step
Performance standard
Test procedures
Equipment
Test
No.
equipment
under test
3
Meter must indicate transcon-
Connect 4k-ohm resistor across
N/A
Same.
d u c t a n c e . Value must be
D-S terminals of transistor
within ±10 percent of full-
under test.
scale range of value indicated
in step 1.
I n s e r t second and third field Transconductance value with
Same.
4
N/A
effect transistor in test jack
resistive load must not vary
by more than 10 percent of
on front panel and determine
full-scale range of unloaded
effect of added resistance per
indication.
steps  2  and  3.
7-22. ICES and IR Measurement Tests
(2) One 500-ohm, 1-percent resistor.
a. Test Equipment and Materials.
b. Test Connections and Conditions. As indi-
cated below.
(1) One good NPN or PNP transistor, the ICES
of which has been checked on a standard transis-
c. Procedure.
tor test set.
Control settings
Step
Performance standard
Test procedures
Test
No.
Equipment
under test
equipment
N/A
1
Meter indicates zero on scale.
Operate METER ZERO to ad-
FUNCTION switch:
I  CES I  R .
just meter zero.
POLARITY  switch:
N P N or PNP to
correspond  with
s e l e c t e d transis-
tor.
R A N G E switch: X
10.
2
N/A
FUNCTION switch:
Zero meter using ICES C O A R S E
Meter indicates zero on scale.
I CES
and FINE controls.
3
N/A
Zero meter using METER
I  CES switch:
press
Meter indicates zero on scale.
ZERO control.
and hold in READ
position.
4
N/A
ICES switch: as indi-
Repeat steps 2 and 3.
Meter indicates zero in both
cated
ZERO and READ positions of
ICES. switch.
5
N/A
R A N G E switch: to
Insert transistor in test jack on
Meter indicates zero with ICES
setting appropri-
front panel. Connect 500-ohm
switch in ZERO position.
ate for device.
resistor across C­E termi-
nals. Use ICES COARSE and
FINE controls to zero meter
6
N/A
I  C E S. switch: press
Check indication on ICES , ICO , IH
Meter indication is within ±10
percent of ICES value deter-
and hold in READ
scale of meter.
position.
mined by standard test set.
M e t e r indication is within
Repeat tests using two other
N/A
7
transistors which require use
of I  CES value  de-
±10percent
termined by standard test
of two remaining ranges.
set.
7-14

Privacy Statement - Press Release - Copyright Information. - Contact Us