Figure 3-6. Effect of DC Bias on the True Rms Voltage of a Square Wave
Up
TM-11-6625-444-14-2 Voltmeter Digital AN/GSM-64C (NSN 6625-01-124-0834) (EIC: KNK) Manual
Next
Standby Operation

TM 11-6625-444-14-2
f. Select the appropriate V RANGE if the ap-
3-10. Remote Programming
proximate value is known; if the voltage is un-
All of the functions and ranges of the voltme-
known or its range will be extensive, select the
ter may be controlled remotely, with the excep-
autoranging mode by pushing the AUTO switch
tion of AC power, RATlO FULL RANGE ADJ.,
to the in Position.
GUARD NORM/EXT, INPUT FRONT/REAR; via
t h e REMOTE PROGRAM INPUT connector
h. Observe the display. It will indicate the
located on the rear panel. Refer to table 3-4 for
measured direct-coupled true rms ac voltage
information regarding remote programming.
combined with the dc level on which it is su-
per-imposed. (See figure 3-6.)
NOTE
A logic IOW (0 to + 0.5 V) at the re-
spective pin of A2J16 actuates the
g. Push the POWER switch to the ON posi-
associated function.
tion.
Table 3-4. Remote Program Connector, A2J16
A2J16
A2J16
Function
Pin No.
Function
Pin No.
Digital COMMON (
)
13
1
HOLD (See note)
2
OHMS
14
2k
RANGE
3
15
AC VOLTS
RESET > 1 usec
16
4
FILTER LP/10 kHz
2 V/ 20 k
RANGE
17
5
AC + DC VOLTS
6
18
SLOW RESPONSE*
20 V/200 k
RANGE
19
7
DC/DC RATlO
20
8
200 V/2 M
RANGE
21
9
AC/DC RANGE
22
10
23
11
1000 V/ 20 M
RANGE
Case GROUND (
)
AC/AC RATIO
24
12
* AC or AC + DC VOLTS mode only
If no RANGE is called; AUTO (Auto-range) is automatically selected.
If no FUNCTION called; DC VOLTS mode is automatically selected.
NOTE
A logic "low" disables read rate gen-
erator and enables remote reset line.
Bringing remote reset line low for at
least 1 usec will produce one sample
reading.
Required 26 pin Remote Program Input Connector, Ballantine P/N: 31-10322-0A and
26 Wire Ribbon Cable, Ballantine P/N: 7 9 - 1 0 1 1 1 - 0 A .