Back
Table 6-14. Buffer DC Alignment
Up
TM-11-6625-444-14-1 Voltmeter Digital AN/GSM-64B (NSN 6625-00-022-7894) (EIC: KPF) Plug-In Electronic Test Equipment PL-1370/GSM-64B Manual
Next
CHAPTER 7. MATERIAL USED IN CONJUNCTION WITH AN/GSM-64B

TM 11-6625-444-14-1
EXTERNAL REFERENCE terminals in 0.1
control (A3R50) until the digital
display changes
to 100.00% ± 50 digits).
VDC increments. At - 10.5 VDC, the voltmeter
(31 ) Reduce the voltage to
the INPUT and
should read 100.000 %, but at -10.6 VDC,
EXTERNAL
REFERENCE
t e r m i n a l s to
should read 100.00% ± 50 digits). If the range
-10.000 VDC. Note that the
digital display
change does not occur between -10.5 VDC and
again indicates 100.000%.
10.6 VDC, repeat steps (27) through (31).
(32) Increase the voltage to
the INPUT and
6-32