![]() ![]() (5) If the meter goes off scale to the left,
reverse the test lead prods.
(6) Read the meter indication on the DC
scale.
d. 0 to 1,000 Volts Ac.
(1) Turn the FUNCTION switch to A C
VOLTS.
(2) Plug the black test lead into the OHMS-
DC±AC jack.
(3) Plug the red test lead into the appropri-
ate 1000 OHMS PER VOLT AC DC jack.
(4) Connect the black test lead prod and the
red test lead prod to the points in the circuit be-
tween which the voltage is to be measured.
(5) Read the meter indication on the A C
scale.
a. Multimeter Zero Adjusting. Zero adjust the
multimeter before making resistance measure-
ments. Each time the range switch is turned to a
different resistance range, repeat the zero ad-
justing procedures as follows:
(1) Turn the FUNCTION switch to OHMS.
NOTE
Do not leave the FUNCTION switch in
the OHMS position any longer than nec-
essary to perform the resistance tests.
Leaving the switch in this position will
drain the batteries.
|
|