Page Title: Procedure for Reading Plate Current (Less Than 50 Milliam-peres) of Triode Tubes
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
Voltage Regulator and Gas Recifier Test
Up
TM-11-6625-316-12 Test Sets Electron Tube TV-2/U TV-2A/U and TV-2B/U Manual
Next
Thyratron Test

c. S e t t h e q u a l i t y S H U N T c o n t r o l t o t h e n u m b e r s p e c i f i e d o n
the tube test data roll chart.
d. Set the FUNCTION switch to the VR position.
e. T u r n t h e S I G N A L - V . R . a n d t h e B I A S f i n e c o n t r o l s t o t h e i r
extreme counterclockwise (zero) positions.
f. D e p r e s s a n d h o l d d o w n t h e P R E S S T O T E S T P 5 s w i t c h .
Note the current through the tube as indicated on the PLATE
meter and note the reading of the PERCENT QUALITY meter.
g. W i t h t h e P 5 s w i t c h h e l d i n i t s o p e r a t e d p o s i t i o n , a d j u s t t h e
S I G N A L - V . R., PLATE, and SCREEN fine controls until the VR-
50 MA scale of the PLATE meter indicates the minimum load
current through the tube as specified on the tube test data roll
chart. Note the reading of the PERCENT QUALITY meter.
h. W i t h t h e P R E S S T O T E S T P 5 s w i t c h s t i l l h e l d i n i t s o p e r -
a t e d position, adjust the SIGNAL-V.R. fine control until the VR­
50 MA scale of the PLATE meter indicates the maximum load
c u r r e n t specified on the chart. Note the reading of the PERCENT
QUALITY  meter.
i. D e t e r m i n e t h e r e g u l a t i n g a b i l i t y o f a t u b e b y t a k i n g t h e
difference between the maximum and minimum load currents
through the tube as read on the PERCENT QUALITY meter.
If the difference between the two readings is greater than the
maximum difference indicated on the tube test data roll chart,
discard the tube.
Note. A quality SHUNT control setting of 50 is specified for many voltage
regulator tubes. When so specified, the reading on the PERCENT QUALITY
meter indicates directly the voltage drop across the tube under test. The
difference between the voltage drop at minimum and maximum load currents
as read on the PERCENT QUALITY meter then is equal to the regulating
ability of the tube in volts.
j. W h e n t h e t e s t i s c o m p l e t e d , s e t t h e O N - O F F p o w e r s w i t c h
to the OFF position, remove the tube under test from the test
socket, and return all switches and controls to their safety posi-
tions (par. 15d).
25. Procedure for Reading Plate Current (Less Than 50 Milliam-
peres) of Triode Tubes
a. P e r f o r m t h e o p e r a t i o n s i n d i c a t e d i n p a r a g r a p h s 1 5 a n d 1 6 .
b. S e t a l l S E L E C T O R S s w i t c h e s a s i n d i c a t e d o n t h e t u b e t e s t
data roll chart with the exception of the PLATE and SCREEN
SELECTORS switches.
c. Set the PLATE selector switch to the 0 position.