Section II. Troubleshooting Procedures
TROUBLESHOOTING
·
Introduction
1. The table lists the common malfunctioning which you may find during the operation or maintenance of the ME-
526A/USM. You should perform the tests/inspections and corrective actions in the order given.
2. This manual cannot list all malfunctions that may occur, nor all tests or inspections and corrective actions. If a
malfunction is not listed or is not corrected by listed corrective action, notify your supervisor.
MALFUNCTION
TEST OR INSPECTION
· CORRECTIVE ACTION
·
1. NO DISPLAY INDICATION, NO DBM METER MOVEMENT
WHEN OFF-MV SWITCH PLACED IN MV POSITION.
Step 1.
Check for ac power.
· Verify ac power present
Step 2.
Check fuse.
· Verify fuse not open.
2. CANNOT ZERO DISPLAY AND DBM METER WITH
ZERO CONTROL.
Step 1. Check that FULL SCALE 1 mV pushbutton
switch is depressed.
· Verify I m V pushbutton Is depressed.
Step 2. Check front panel PROBE connector for
proper connection of RF PROBE.
· Verify PROBE connector is secure.
Step 3. Check that there is no rf input and the 50-
OHM ADAPTER is attached to the RF
PROBE.
· Verify input removed. Attach 50-OHM ADAPTER.
3. INDICATOR DISPLAY IS KNOWN TO BE INCORRECT.
Step 1. Check connection made to signal under test.
CAUTION
When the RF PROBE must be replaced send the ME-
526A/USM to higher level maintenance. The ME-
526A/USM must be calibrated with the RF PROBE
with which it is used.
Step 2.
Check RF PROBE tip (if used) for
proper connection to RF PROBE.
· Verify RF PROBE tip makes connection to RF PROBE.
Replace tip if faulty. If RF PROBE is faulty, higher level
maintenance required.