CHAPTER 1. INTRODUCTION
Up
TM-11-6625-298-14 Ohmmeter ZM-21/U ( NSN 5950-00-645-2191 ) Ohmmeter ZM-21A/U ( N S N 6625-00-643-1030) and Ohmmeter ZM-21B/U ( NSN 6625-00-581-2466) Manual
Next
CHAPTER 2. SERVICE UPON RECEIPT AND INSTALLATION

TM 11-6625-298-14
Figure 1-2. Ohmmeter ZM-21(*)/U, exterior view.
1-7. Purpose and Use
1-8. Tabulated Data
Resistance range . . . . . . . . 0 through 1,000 megohms.
a. Ohmmeter (fig. 1-2) is used to determine the
Test potential..................... 500 volts dc ±5 percent (when
insulation condition of telephone cables, of trans-
measuring values below 2
formers, between windings and ground of rotating
megohms, the voltage drops
equipment, and of all other types of electrical
materially).
equipment for which insulation resistance is an
Accuracy . . . . . . . . .. Within 1 percent of any car-
dinal calibration point.
important factor.
NOTE
b. The ohmmeter applies a high potential to
Insulation resistance above 1,000 megohms
equipment under test in order to detect low in-
sulation  resistance which may not indicate
is indicated by a point marked infinity
satisfactorily on a low potential ohmmeter.
on the meter scale.
c. Regular use of the ohmmeter can minimize
1-9. Items Comprising an Operable Ohmmeter ZM-
failure in circuits or equipment caused by faulty
21(*)/U
insulation.  A sudden lowering of insulation
Refer to table 1-1 for items comprising an operable
resistance indicates a fault that should be in-
Ohmmeter ZM-21(*)/U.
vestigated immediately.