SCOPE - cont
Up
TM-10-6640-222-13P Emcee Micro-Separometer Mark V Deluxe Model No. 1140 Manual
Next
JET A TEST PROCEDURE

MICRO-SEPAROMETER MARK V DELUXE OPERATION MANUAL Test performed outside of these temperature limits may not meet ASTM repeatability/reproducibility criteria. 7.2 Remove the instrument from the protective foam case.  Open the case and remove the Six-Pack box from the lid. Insert  your  finger  into  the  hole  in  the  right  panel  and  raise  the  panel  until  completely  vertical  and  locked  in  place.    If power is available, check to be certain the source voltage matches the power cord label and connect the power cord. 7.3 Have ready a supply of syringes, ALUMICELÒ  COALESCERS, vials, plugs, pipette tips, drip pan, and a container for clean water.  All but the water container are provided with the instrument and Six-Pack. 7.4 The syringe drive speeds are pre-set at the factory and are automatically checked during each test.  The syringe drive time is measured and compared to an internal clock.  As soon as the syringe reaches the bottom limit the "SYR.  " alert indicator will illuminate if the syringe speed is out of limits.  The syringe alert lamp will remain on until the end of the test to alert the operator the data may be suspect.  If the syringe speed is further out of limits the audible alarm will sound 3 times and any test data should be discarded. 7.5 If  the  instrument  is  to  be  operated  in  a  location  where  an  AC  outlet  is  not  available,  the  instrument  should  be charged  for  approximately  16  hours  before  testing.    The  unit  will  function  properly  for  approximately  25  tests  before battery recharging is necessary. 8.0 PREPARATION OF SAMPLE 8.1 Obtain and handle the samples with the utmost care and cleanliness as required in the ASTM Procedure D-3948. Before pouring the test sample from container, wipe the container outlet thoroughly with a clean, lintless wipe.  Then pour the test sample into a clean beaker or directly into the syringe. 8.2 If the sample for test is not within the test temperature limits let the container stand or preferably put sample in a clean beaker and let it come to operating temperature. COPYRIGHT 1984  EMCEE ELECTRONICS, INC. REVISION DATE:  FEBRUARY 12, 1987 PAGE 6

Privacy Statement - Press Release - Copyright Information. - Contact Us - Support Integrated Publishing