CLEAR AND BRIGHT - cont
Up
TM-10-6640-222-13P Emcee Micro-Separometer Mark V Deluxe Model No. 1140 Manual
Next
DIESEL TEST PROCEDURE - cont

MICRO-SEPAROMETER MARK V DELUXE OPERATION MANUAL 12.0 DIESEL TEST PROCEDURE 12.1 Prepare the DELUXE MICRO-SEPAROMETER as stated in section 7.0. 12.2 Momentarily  depress  the  "ON"  switch.    The  annunciator  lamps  located  on  switches  A  through  G  in  the  TEST SELECT section will then be scanning for your selection.  Depress switch "D".  This will initiate the standard automatic program for Diesel fuel. 12.3 The  annunciator  lamp  on  the  "START"  switch  will  be  illuminated  which  indicates  the  program  can  be  started  by momentarily depressing this switch. 12.4 Remove  a  plunger  from  a  50  ml.    syringe,  insert  a  plug  in  the  bottom,  add  approximately  50  ml.    of  fuel  to  be tested, place the syringe on the emulsifier turning to lock in place.  Caution should be exercised to center the emulsifier blade   in   the   syringe   (Figure   5).      Press   the   "START"   pushbutton   which   will   automatically   clean   the   syringe   for approximately 30 seconds and also raise the syringe drive to its UP position. 12.5 Add about 15 or 20 ml.  of the fuel to be tested to a clean vial and insert into the TURBIDIMETER (the hole directly above the PROGRAM and METER sections).  Align the black vial mark with the white line on the front panel. 12.6 Remove the syringe from the emulsifier, empty the fuel, draining thoroughly, and add exactly 50 ml.  of test fuel. 12.7 Using a fresh tip on the hand pipette, add 50 microliters of clean water to fuel sample as follows.  Push in plunger, immerse tip just below the water surface, release the plunger and withdraw from the water slowly to avoid water drips adhering to the outside of the tip.  Immerse the tip of the pipette just below the fuel surface in the center of the syringe, to ensure  the  water  drops  break  away  cleanly  and  fall  to  the  bottom;  push  the  plunger,  withdraw  the  pipette,  and  then release the plunger. 12.8 Place the syringe on the emulsifier, turning to lock in place.  Caution should be exercised to center the emulsifier blade in the syringe (Figure 5). COPYRIGHT 1984 EMCEE ELECTRONICS, INC. REVISION DATE:  FEBRUARY 12, 1987 PAGE 13

Privacy Statement - Press Release - Copyright Information. - Contact Us - Support Integrated Publishing