Click here to make tpub.com your Home Page

Page Title: Aircraft Indicator
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
Resistance Check
Up
TB-9-4920-454-35 Tester Exhaust Gas Temperature Howell Instruments Inc. Model BH112JB-( ) Manual
Next
Final Procedure

TB 9-4920-454-35
(6) Adjust resistance standard for a 0 indication on RESISTANCE meter.  If
resistance standard does not indicate between 21.95 Ω and 22.05 Ω, perform b (3) below.
(7) Repeat (4) and (6) above for 15 Ω.  If resistance standard does not indicate
between 14.95 Ω and 15.05 Ω, perform b (4) below.
(8) Repeat (4) and (6) above for 8 Ω.  If resistance standard does not indicate
between 7.95 Ω and 8.05 Ω, perform b (5) below.
(9) Repeat (4) and (6) above for 4.21 Ω. If resistance standard does not indicate
between 4.16 Ω and 4.26 Ω, perform b (6) below.
b. Adjustments
(1) Connect multimeter to RESISTANCE CHECK meter terminals, observing
polarity. Adjust resistance standard to 22 Ω. Adjust R13 (fig. 12) ccw. Record multimeter
indication, disregarding polarity sign.  Adjust R13 (fig. 12) cw.  Record multimeter
indication, disregarding polarity sign.
(2) Determine the mean value of indications recorded in (1) above. Adjust R2 (fig. 12)
to this value, then disconnect multimeter.
(3) Set resistance standard to 22 Ω and adjust R13 (fig. 12) for a 0 indication on
RESISTANCE CHECK meter (R).
(4) Set resistance standard to 15 Ω and adjust R10 for a 0 indication on
RESISTANCE CHECK meter (R).
(5) Set resistance standard to 8 Ω and adjust R7 for a 0 indication on
RESISTANCE CHECK meter (R).
(6) Set resistance standard to 4.21 Ω and adjust R4 for a 0 indication on
RESISTANCE CHECK meter (R).
15. Aircraft Indicator
a. Performance Check
(1) Replace BH-823 adapter with BH-822 adapter in equipment setup shown in
figure 11, and remove resistance standard connection.
(2) Set FUNCTION SELECT switch to A/C IND CHK and RESISTANCE & A/C
INDICATOR CHECK switch to NULL BAL.

Privacy Statement - Press Release - Copyright Information. - Contact Us