Heater Probe Control
Up
TB-9-4920-454-35 Tester Exhaust Gas Temperature Howell Instruments Inc. Model BH112JB-( ) Manual
Next
Heater Probe Control-cont.

TB 9-4920-454-35
(b) T/C INPUT to HEATER CABLE.
(c) STD DAY to OFF.
(d) MASTER POWER switch to OFF.
(3) Remove power from TI by disconnecting power cable from 115 V ac power
source.
WARNING
115 V ac potential exists between pin A and/or B of the
HEATER CABLE connector and ground when MASTER
POWER switch is in the OFF position and when power cable
is connected to ac source.
DEATH ON CONTACT and
destruction of calibrator may result if personnel fail to observe
safety precautions.
(4) Connect equipment as shown in figure 8, CONNECTION A. Set multimeter for
ac voltage measurements.
Figure 8. Heater probe.
(5) Reconnect power cable to 115 V ac source and set MASTER POWER switch to ON.
14