![]() ![]() TB 9-4920-454-35
(5) Refer to table 3 and adjust calibrator to millivolt value corresponding to ambient
temperature recorded in (1) above.
(6) Multimeter will indicate between 7.255 and 7.295 mV dc; if not, perform b (l)
below.
(7) Replace jumper at TP2 (fig. 4) and reconnect J14 to P14.
(8) Set TEMPERATURE switch to CAL 600 oC. If TEMP oC indicator does not
display 600 perform b (2) below.
b. Adjustments
(1) Adjust R6 (fig. 4) for 7.275 mV dc indication on multimeter (R).
10. % RPM Indicator
a. Performance Check
(1) Set RPM switch to CAL 100% RPM and adjust RPM CAL ADJ control for
100.0 indication on % RPM indicator.
(2) Set RPM switch to N1.
(3) Short pins A and B on RPM INSTRUMENT CABLE receptacle, using test
probe hooks.
(4) If % RPM indicator does not indicate .0 or .1, perform b (l) through (3) below.
(5) Set RPM switch to N2.
(6) Remove shorting connection from pin A, and connect to pin C on RPM INPUT
INSTRUMENT CABLE receptacle.
(7) %RPM indicator will indicate .0 or .1.
(8) Remove short from pins B and C and connect equipment as shown in figure 5.
|
|