TB-9-4920-454-35 Tester Exhaust Gas Temperature Howell Instruments Inc. Model BH112JB-( ) Manual
Next
Figure 2. Temperature indicator check.

TB 9-4920-454-35
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUTS(S)  to minimum after each step within the
performance check where applicable.
a. If required to make adjustments, remove TI from case and install probe controller on
mounting bracket shown in figure 1.
b. Position switches as listed in (1) through (3) below:
(1) FUNCTION SELECTOR to OFF.
(2) STD DAY to OFF.
(3) MASTER POWER to OFF.
c. Connect BH-499A power cable, supplied with TI, to power input on TI and to 115 V ac, 50 to
400 Hz power source.
d. Connect BH15184A1 cable, supplied with TI, to power interconnectors.
e. Set MASTER POWER switch to ON and allow 15 minutes for warm-up.
f. Adjust BRIGHT DISPLAY control for suitable intensity of LED readouts.
9"
7 1/2"
7"
MATERIEL 1" X 4" WOOD
Figure 1. Probe controller support.