SECTION I IDENTIFICATION AND DESCRIPTION
Up
TB-9-4920-454-35 Tester Exhaust Gas Temperature Howell Instruments Inc. Model BH112JB-( ) Manual
Next
Equipment Setup

TB 9-4920-454-35
5. Accessories Required.  The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Range: 0 mV to 10 V dc
Fluke, Model 5720A (5700A/EP)
Accuracy:  ±.003%
(p/o MIS-35947)
Range: 20 mV, 70 Hz
Accuracy:  ±3.0%
FREQUENCY COUNTER
Range: 103 ms to 15.873 ms
Fluke, Model PM6681/656
Accuracy:  ±.033%
(PM6681/656)
MULTIMETER
Range: 0 to 1000 V dc
Hewlett-Packard, Model 3458A
Accuracy:  ±.07%
(3458A)
OSCILLATOR
Range: 1 V ac
Tektronix, Type SG505
Frequency: 63 to 77 Hz
(MIS-30526/11)
Accuracy:  ±.25%
OSCILLOSCOPE
Tektronix, Type 2465B-46
Range: 14 to 100 µs
(2465B-46)
Accuracy:  ±3.0%
RESISTANCE STANDARD
Range: 4.21 Ω to 22 Ω
Biddle Gray, Model 71-650
(71-650)
Accuracy:  ±0.03%
Range: 10 kΩ and 100 kΩ
Accuracy:  +2.5%
THERMOMETER
Range: 70 to 85 oF
Azonix, Model A1012 (MIS38958)
Accuracy: ±1.0
oF
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in the manufacturer's manual and TM 55-4920-401-
13&P for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.