![]() ![]() TB 9-4920-452-35
(2) Repeat technique in (1) above to obtain a 10, 20, 30, 40, 50, 60, 70, and 80
percent cardinal point on TI. Indications as observed on TI will be within limits specified in
applicable table 3 or 4.
(3) Repeat technique in (1) and (2) above for remaining TI gage to be calibrated.
Indications as observed on TI will be within limits specified in applicable table 3 or 4.
b. Adjustments. No adjustments can be made.
9. Frequency Meter
a. Performance Check
(1) Connect equipment as shown in figure 1.
Figure 1. Frequency meter - test equipment hookup.
(2) Set frequency counter CHANNEL A 50Ω/1MΩ switch to 1MΩ and CHECK/
COMA/SEP switch to SEP.