Figure 3. Consolidated Airborne Systems, Models TF20-1 and TF579 - adjustment locations.
Up
TB-9-4920-451-35 Fuel Quantity Gage Test Set Simmons Precision Inc. Model 472090-002 Gull Airborne Instruments Model GTF-4 Manual
Next
Final Procedure

TB 9-4920-451-35
j.  Adjust autotransformer between 105 and 125 V. Capacitance meter will indicate
within ±5 percent of reading stamped on front panel.
20. Capacitance
a. Performance Check
(1) Connect variable capacitor, 1150 to tank unit A coax connector using shielded
test probe (p/o TI).
(2) Connect variable capacitor GND terminal to TI B-UNSH terminal using
unshielded test probe (p/o TI).
(3) Set variable capacitor for 100 pF.
(4) Set FUNCTION SELECTOR switch to TANK UNIT TEST-UNSH and set
RANGE SELECTOR switch to X1. If capacitance indicator meter does not indicate
between 99.50 and 100.50 pF, perform b below.
(5) Repeat (4) above using settings, indications, and adjustments listed in table 9.
Table 9. Capacitance
Test instrument
Variable
Indications
capacitor
(pF)
Adjustments
(pF)
Range
Min
Max
300
X3
99.50
100.5
C103
0
1000
X10
99.50
100.5
C105
0
1000
X50
19.50
20.50
---
50001
X50
99.5
100.5
---
1Subitute
capacitance standard.
b. Adjustments. Adjust C101 (fig. 3) for a capacitance indication of 100 pF.
21. Resistance
a. Performance Check
(1) Set RANGE SELECT switch to ZERO CAL. If megohmmeter does not indicate
zero adjust ZERO ADJ (front panel) for zero indication.
(2) Set CAP RES-CHECK switch to MEGOHMS AC position and set A-B
MEGOHMMETER and RANGE SELECTOR switch to MIDSCALE CAL position. If
megohmmeter does not indicate midscale, adjust MIDSCALE ADJ (front panel) for
midscale indication.
(3) Connect resistance standard (NO.1) to TANK UNIT A and TANK UNIT B
using cables furnished with TI.