Capacitance Substitution Calibration
Up
TB-9-4920-451-35 Fuel Quantity Gage Test Set Simmons Precision Inc. Model 472090-002 Gull Airborne Instruments Model GTF-4 Manual
Next
Table 6. Resistance

TB 9-4920-451-35
(14) Remove cable from TO IND LOZ and shorting plug from TO IND LOZ COMP.
(15) Connect shorting plug (p/o TI) to TO IND LOZ and cable to TO IND LOZ
COMP.
(16) Position controls as listed in (a) through (c) below:
(a) MODE SELECT switch to MEAS INT.
(b) SIMULATOR COMPENSATOR SET control for 40 pF on digital display.
(c) MODE SELECT switch to SIM.
(17) Precision component analyzer will indicate between 39.8 and 40.2 pF.
(18) Repeat (16) and (17) for each value through 100 pF as shown in table 5.
Table 5. Capacitance Substitution
Test
Precision component analyzer
instrument
indications (pF)
indications
Min
Max
60
59.7
60.3
80
79.6
80.4
100
99.5
100.5
180
179.1
180.9
300
298.5
301.5
400
398.0
402.0
600
597.0
603.0
800
796.0
804.0
(19) Set SIMULATOR COMPENSATOR 100 pF switch to IN.
(20) Repeat (16) and (17) for 180 pF value.
11. Resistance Measurement Calibration
a. Performance Check
(1) Set switches as listed in (a) through (c) below:
(a) MODE SELECT to MEG.
(b) MEGOHMMETER SELECT to EXT.
(c) S8 switch (fig. 1) to IN.
(2) If MEGOHMS meter does not indicate value stamped above S8 switch (fig. 1),
perform b(1) below.
(3) Set S8 switch (fig. 1) to OUT position and S9 switch (fig. 1) to IN.
(4) If MEGOHMS meter does not indicate value stamped above S9 switch (fig. 1),
perform b(2) below.
10 CHANGE 1