Figure 1. Simmons Precision, Inc. Model 472090-002, controls and adjustments locations.
Up
TB-9-4920-451-35 Fuel Quantity Gage Test Set Simmons Precision Inc. Model 472090-002 Gull Airborne Instruments Model GTF-4 Manual
Next
Capacitance Measurement Calibration

TB 9-4920-451-35
(3) Set S5 (left most toggle switch, fig.1) to the IN position.
(4) If digital display does not indicate value stamped above S5 (fig.1), perform b (1)
below.
(5) Set switches as listed in (a) through (c) below (fig. 1).
(a) S5 to OUT position.
(b) S6 to IN position. If digital display does not indicate value stamped above
S6 perform b (2) below.
(c) S6 to OUT position.
(6) Repeat (3) through (5) (c) until both display values are equal to value stamped
above each switch and no further adjustments are required.
(7) Set switches as listed in (a) through (g) below (fig 1):
(a) RANGE to HI position.
(b) S6 to IN position. If display does not indicate value stamped above S6,
perform b (3) below.
(c) S6 to OUT position and S7 to IN position. If display does not indicate value
stamped above S7 , perform b (4) below.
(d) S7 to OUT position.
(e) MODE SELECT to MEG position and MEGOHMMETER SELECT to
EXT position.
(f) S8 to IN position. If MEGOHMS meter does not read value stamped above
S8 , perform b (5) below.
(g) S8 to OUT and S9 (farthest right toggle switch fig.1) to IN positions.
(8) If MEGOHMS meter does not read value stamped above S9 (fig 1), perform b (6)
below.
(9) Set S9 switch (fig.1) to OUT position.
b. Adjustments
(1) Adjust LO RANGE OFFSET trimmer screw (fig. 1) until display reads same as
value stamped above S5 switch (fig.1) (R).
(2) Adjust LO RANGE GAIN trimmer screw until display reads same as value
stamped above S6 switch (fig.1) (R).
(3) Adjust HI RANGE OFFSET trimmer screw until display reads same as value
stamped above S6 switch (fig.1) (R).