SECTION II EQUIPMENT REQUIREMENTS
Up
TB-9-4920-451-35 Fuel Quantity Gage Test Set Simmons Precision Inc. Model 472090-002 Gull Airborne Instruments Model GTF-4 Manual
Next
Equipment Setup

TB 9-4920-451-35
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CAPACITANCE
Range: 2000 and 5000 Pf
Arco, Model SS32 (7907233)
STANDARD
Accuracy:  +0.1%
Wayne-Kerr, Model 6425B
PRECISION
Capacitance-
COMPONENT
Range: 0.1 pF to 2000 µf
ANALYZER
Accuracy: ± 0.05 %
Range: 0.01Ω to 1,111,111
RESISTANCE
Biddle-Gray Model
Accuracy: ±0.03% of indication
STANDARD (NO.1)
71-631 (7910328)
RESISTANCE
ICC, Model CR10M
Range: 100 KΩ to 10 MΩ
STANDARD (NO.2)
(8598965)
Accuracy: ±0.02%
RESISTANCE
ICC, Model CR100M
Range: 10 to 100 MΩ
STANDARD (NO.3)
(8598966)
Accuracy: ±0.05%
RESISTANCE
ICC Model CR1000M
Range: 1 to 1000 MΩ
STANDARD (NO.4)
Accuracy: ±1%
(6579478)
RESISTANCE
Penn-Airborne, Model
Range: 10,000 MΩ
STANDARD (NO.5)
Accuracy: ±3.5%
9A-5120-103 (MIS-10412-4)
VARIABLE
Range: 40 to 1000 pF
General Radio, Model 1422D
CAPACITOR
Accuracy: +0.1%
(8579475)
1
1100
to 1000 pF range: +0.6 pF.
CALIBRATION PROCESS FOR SIMMONS PRECISION, INC.
MODEL 472090-002
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in the manufacturers' manuals for this TI.
d. Unless otherwise specified all controls and control settings refer to the TI.