Page Title: SECTION II EQUIPMENT REQUIREMENTS
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
SECTION I IDENTIFICATION AND DESCRIPTION
Up
TB-9-4920-451-35 Fuel Quantity Gage Test Set Simmons Precision Inc. Model 472090-002 Gull Airborne Instruments Model GTF-4 Manual
Next
SECTION III CALIBRATION PROCESS FOR SIMMONS PRECISION, INC. MODEL 472090-002

TB 9-4920-451-35
Table 1. Calibration Description
Test instrument parameters
Performance specifications
Simmons Precision, Inc., Model 472090-002
Capacitance measurement
Range: 40 to 999.9 pF
(external)
Accuracy: ±0.5% of reading
Range: 1000 to 2000 pF
Accuracy: ± .5% of reading
Capacitance substitution
Range: 40 to 999.9 pF
(internal)
Accuracy: +0.5% of reading
Range: 1000 to 2000 pF
Accuracy: ± .5% of reading
Resistance
Range: 1.0 to 1000 MΩ
Accuracy: +l increment
-.4 increment
Gull Airborne Instruments, Model GTF-4
Capacitance
Range: 0 to 8000 pF in 4 ranges1
Accuracy: ±0.5% of range setting
Resistance
Range: 0 to 100,000 MΩ in 4 ranges
Accuracy: 1.0 to 1000 mΩ +.5% of reading
1000 to 10,000 MΩ +10% of reading
10,000 to 100,000 MΩ +15% of reading
0.1 to 1.0 MΩ +1.0% of range
Consolidated Airborne Systems, Models TF20-1 and TF579
Capacitance
Range: 0 to 5000 pF
Accuracy: +0.5% of FS
Resistance
Range: 0 to 10,000 MΩ
Accuracy: 10% of reading
1Verified
from 35 to 2000 pF only.
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity when
the equipment listed in table 2 is not available. The items selected must be verified to
perform satisfactorily prior to use and must bear evidence of current calibration.  The
equipment must meet or exceed the minimum use specifications listed in table 2.
5. Accessories Required.  The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.