Page Title: Insulation Resistance Check
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
Resistance Check
Up
TB-9-4920-364-35 Exhaust Gas Temperature Tester Howell Instrument Inc. Model BH112JA36 Manual
Next
Rpm Check

TB 9-4920-364-35
Figure 2. Adjustment locations -bottom view.
(9) Repeat technique of (4) through (7) above for each setting of RESISTANCE
CHECK SW-3 switch and set resistance standard to the value indicated on SW-3. TI
GALVO-1 will indicate within ±5 divisions at all SW-3 settings.
b. Adjustments. Adjust RES CHECK TRIMPOT (fig. 2) for 0 indication on GALVO-1 (R).
10. Insulation Resistance Check
a. Performance Check
(1) Adjust resistance standard for 0 ohms.
(2) Connect TI INPUT RES. INSUL. P-2 to HIGH and LOW terminals of
resistance standard using cables and insulation check adapter (BH485 and BH821, p/o TI).
(3) Set SELECTOR SWITCH SW-1 to INSUL. and SW-8 switch to RX100. If
INSULATION CHECK meter does not indicate 0, perform b (1) below.
(4) Adjust resistance standard for indication of 10,000 ohms on INSULATION
CHECK meter. If resistance standard does not indicate between 9000 and 11,000 ohms,
perform b (2) below.