![]() ![]() TB 9-4920-363-24
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.
Manufacturer and model
Common name
Minimum use specifications
(part number)
PNEUMATIC PRESSURE
Range: 0 to 120 inH2O
Druck, Model DPI145/R
STANDARD
Accuracy: ±0.1% of FS pressure
(MIS-45842)
Range: 0 to 24.5 psi
0.6 to 75.6 inHg abs
Accuracy: ±0.05 psi,
±0.033 inHg
CALIBRATION PROCESS
6. Preliminary Instructions
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
7. Equipment Setup
a. Place TI on a level platform or bench.
b. Remove protective cover from TI.
c. Position TI controls as listed in (1) through (5) below:
(1) VACUUM selector control valve to OFF.
(2) PRESSURE selector control valve to OFF.
(3) VACUUM and PRESSURE INCREASE and DECREASE control valves fully closed.
(4) POWER switch to OFF.
|
|