Click here to make tpub.com your Home Page

Page Title: SECTION III CALIBRATION PROCESS
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
SECTION I IDENTIFICATION AND DESCRIPTION
Up
TB-9-4920-362-35 Test Box Assy Pilt Assist/Nulling Sirkorsky Model 70700-20678-041 Manual
Next
Equipment Setup

TB 9-4920-362-35
5. Accessories Required. The accessories listed in table 3 are issued as indicated in
paragraph 4 above and are used in this calibration procedure. When necessary, these items
may be substituted by equivalent items, unless specifically prohibited.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model (part
Item
Common name
Minimum use specifications
number)
A1
AUTOTRANSFORMER
Range: 115 V ac
General Radio, Model
W10MT3AS3 (7910809)
Accuracy: ±1%
A2
DC POWER SUPPLY
Range: 28 V dc
NJE, Model CS36CR30D2
(7907346-2)
A3
DIGITAL MULTIMETER
Range: -10.5 to +10.5 mA
Tektronix, Type DM501A
(DM501A)
Accuracy: ±1.25%
A4
DIGITAL VOLTMETER
Range: -15.5 to +28.5 V dc
Hewlett-Packard, Model
349AOPT060 (3490AOPT060)
Accuracy: ±0.45%
Table 3. Accessories Required
Item
Common name
Description (part number)
B1
ADAPTER
Single banana jack to pin plug (black) (7907528)
B2
ADAPTER
Single banana jack to pin plug (red) (7907517)
B3
LEAD
24-in., No. 18; single banana plug terminations (black) (7907498)
B4
LEAD
24-in., No. 18; single banana plug terminations (red) (7907497)
B5
LEAD1
Miniature pin jack with 12-in., No. 16 wire attached (fabricated locally)
B6
LEAD2
Pin jack to single banana plug (7921032)
1Two
required.
2Three
required.
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name and item identification number as listed in tables 2 and 3. For the identification of
equipment referenced by item numbers prefixed with A, see table 2, and for prefix B, see
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
3

Privacy Statement - Press Release - Copyright Information. - Contact Us