![]() ![]() TB 9-4920-361-24
d. Adjust autotransformer output for 115 V.
e. Connect multimeter to dc power supply and adjust dc power supply for a 28 V
output indication on multimeter.
f. Disconnect multimeter from equipment setup.
g. Set ON-OFF-SIM ONLY switch to ON and allow at least 30 minutes for warm-up.
h. Set ON-OFF-SIM ONLY switch to OFF.
CAUTION
To prevent damage to TI, always turn power off before making
connections.
8. Output Voltages
a. Performance Check
(1) Connect multimeter between TI connector No. 1 pins FF (HI) and HH (LO).
Refer to figure 2 for pin locations.
(2) Set ON-OFF-SIM ONLY switch to SIM ONLY and INTEGRAL LIGHTING switch to
INT. Multimeter will indicate between 4.0 and 6.0 V ac.
(3) Set ON-OFF-SIM ONLY switch to OFF.
(4) Connect multimeter between connector No. 1 pins P (HI) and EE (LO), refer to figure 2.
(5) Set ON-OFF-SIM ONLY switch to ON.
If multimeter does not indicate
between 11.5 and 12.5 V dc, perform b below.
(6) Set ON-OFF-SIM ONLY switch to OFF.
|
|