| ||||||||||
|
|
![]() ![]() TB 9-4920-360-35
Minimum use
Manufacturer and model
Common name
specifications
(part number)
AUTOTRANSFORMER Range: 105 to 125 V ac
General Radio, Model
1%
W10MT3AS3 (7910809)
Accuracy:
FREQUENCY
Range: 250 to 5531 Hz
Fluke, Model PM6681/656
0.05%
COUNTER
(PM6681/656)
Accuracy:
MULTIMETER
Range: 16.140 to 33.523 mV dc
Hewlett-Packard, Model 3458A
1.5%
(3458A)
Accuracy:
Range: 15 to 30 C
THERMOMETER
zonix, Model A1012
A IS38958)
(M
1.5
Accuracy:
CALIBRATION PROCESS
6. Preliminary Instructions
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration. DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Position controls as listed in (1) through (4) below:
(1) POWER switch to OFF.
(2) FUNCTION SWITCH to 0%.
(3) VAR FREQ ADJ switch to 0.
3
|
|
Privacy Statement - Press Release - Copyright Information. - Contact Us |