IDENTIFICATION AND DESCRIPTION
This bulletin provides instructions for the
calibration of Temperature-Speed Simulator Test Set, Howell Model H296A-1.
Manufacturer's manual was used as the prime data source in compiling these instructions.
The equipment being calibrated will be referred to as the TI (test instrument) throughout
this procedure.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Power input requirement
Range: 105 to 125 V ac, 50 to 400 Hz
Range: 400 to 800 C
Temperature
6C
Accuracy:
Percent rpm
Range: 5 to 110% rpm (239.52 to 5531.04 Hz)
0.2%
Accuracy:
EQUIPMENT REQUIREMENTS
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286 or AN/GSM-705.
Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior to
use and must bear evidence of current calibration. The equipment must meet or exceed the
to-one ratio between the standard and TI. Where the four-to-one ratio cannot be met, the
actual accuracy of the equipment selected is shown in parenthesis.
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.