(2) Position TI controls as listed in (a) through (d) below:
TEST SELECT switch to A∅GEN CT.
(a)
(b)
CT CUR READ switch to up.
(c)
GEN/PMG VOLTS RANGE switch to T1.
400 ∼ GEN T1 to in position.
(d)
(3) Adjust resistance standard to 3300 ohms.
(4) Set calibrator to 1 V, 400 Hz and press EX SENS. Set calibrator to OPR.