![]() ![]() TB 9-4920-359-35
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Underfrequency circuit:
Low frequency
Range:
1095 Hz
±10 Hz
Accuracy:
High frequency
Range:
1155 Hz
±10 Hz
Accuracy:
EQUIPMENT REQUIREMENTS
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Sets AN/GSM-287 or AN/GSM-705. Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior to
use and must bear evidence of current calibration. The equipment must meet or exceed the
to-one ratio between the standard and TI. Where the four-to-one ratio cannot be met, the
actual accuracy of the equipment selected is shown in parenthesis.
usage accessories issued as indicated in paragraph 4 above and are not listed in this
calibration procedure. The following peculiar accessories are also required for this
calibration: Power Supply, Sorensen Model DCS 40-30; and Power Amplifier, CPD
Engineering Model 97717.
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Range: 1 V ac
Fluke, Model 5720A (5700A/EP)
at 396 to 1200 Hz
(p/o MIS-35947); w amplifier,
30 V dc
Fluke 5725A/AR (5725A/AR)
Accuracy:
ac ±0.4%, dc ±1.0%
Freq Accuracy: ±0.25%
Range: 0 to 5 A dc
Accuracy: ±0.05 A dc
FREQUENCY
Range: 100 ms to 5.5 s;
Fluke, Model PM6681/656
COUNTER
(PM6681/656)
Accuracy: ±0.02%
MULTIMETER
Range: 8 mV to 28 V dc;
Hewlett-Packard, Model 3458A
10 to 30.4 V ac
(3458A)
Accuracy: ±1% dc; ±0.33% ac
Frequency: 1085 to 1165 Hz
Accuracy: ±0.23%
RESISTANCE
Biddle-Gray, Model 71-631
Range: 1.1 to 3.3 kΩ
STANDARD
(7910328)
Accuracy: ±0.75%
|
|