18. Power Supply
a. Performance Check
NOTE
Do not perform power supply check if all other parameters are
within tolerance.
chassis. Multimeter indication will be between -50 and +50 mV.
(2) Move multimeter + (positive) lead to TI TP3 (fig. 2). Multimeter indication will
be between 4.75 and 5.25 Vdc.
(3) Move multimeter + (positive) lead to TI TP2 (fig. 2). Multimeter indication will
be between 11.5 and 12.5 Vdc.
b. Adjustments. No adjustments can be made.
19. Final Procedure
a. Deenergize and disconnect all equipment and reinstall protective covers on TI.
b. Annotate and affix DA label/form in accordance with TB 750-25.