Click here to make tpub.com your Home Page

Page Title: Figure 10. Dc and ac current - equipment setup
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
Resistance
Up
TB-9-4910-556-35 Calibration Procedure for STE-M1/FVS Test Set Manual
Next
Final Procedure

TB 9-4910-556-35
Figure 10. Dc and ac current - equipment setup .
(2) Press and release STEP key on SC. SC will display:
DC CURRENT
Range: 0 to +2 A dc
Then: ±0.000 A dc
(3) Set calibrator for 1.5 A dc output.
(4) Set calibrator to OPERATE mode.
(5) Set communicator will indicate between 1.46 and 1.53 A dc.
(6) Set calibrator to STANDBY mode.
b. Adjustments. No adjustments can be made.
30. Ac Current
a. Performance Check
(1) Connect equipment as shown in figure 10.
(2) Press and release STEP key on SC. SC will display:
AC CURRENT
Range: 0 to 1.14 A rms
Then: ±0.000 A ac.
(3) Set calibrator for 1 A ac, 400 Hz output.
CHANGE 1 25