100 Millivolt Channel Check
Up
TB-9-4910-556-35 Calibration Procedure for STE-M1/FVS Test Set Manual
Next
SECTION V CALIBRATION PROCESS FOR SET COMMUNICATOR (SC)

TB 9-4910-556-35
(2) Repeat technique of (1) above with + (positive) lead from digital multimeter
connected to CVA terminals N, LL, H, P, and MM. Indication on digital multimeter will
remain between +9.15 and +9.85 V.
(3) Using DATA ENTRY keys, enter 66666 on set communicator and repeat (1) and
(2) above. Digital multimeter indications will remain between -9.15 and -9.85 V.
(4) Press set communicator STOP and CLEAR keys.  Enter 66664 on set
communicator and press GO key.
(5) Press set communicator STOP and CLEAR keys.
b. Adjustments. No adjustments can be made.
23. Frequency
a. Performance Check
(1) Connect equipment as shown in figure 9.
Figure 9. Frequency accuracy ­ equipment setup.

Privacy Statement - Press Release - Copyright Information. - Contact Us