SECTION IV CALIBRATION PROCESS FOR CONTROLLABLE INTERFACE BOX (CIB)
Up
TB-9-4910-556-35 Calibration Procedure for STE-M1/FVS Test Set Manual
Next
Differential Voltage

TB 9-4910-556-35
17. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance checks where applicable.
a. Connect equipment as shown in CONNECTION A (fig. 8).
Figure 8. 10 V and 100 mV channel accuracy - equipment accuracy.
b. Set calibrator to STANDBY and 24 V output.
c. Set CIB PWR switch to ON.
d. Press VTM power switch to ON.
e. Set TEST SELECT switches on VTM to 00.
NOTE
If set communicator is in MULTIMETER MODE, press
pushbutton on lower left of set communicator.