Page Title: Figure 3. Ac current - equipment setup.
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
Ac Current
Up
TB-9-4910-556-35 Calibration Procedure for STE-M1/FVS Test Set Manual
Next
Ac Voltage and Frequency

TB 9-4910-556-35
Figure 3. Ac current - equipment setup.
(2) Set TEST SELECT switches to 95.
(3) Press and hold TEST pushbutton until CAL appears on VTM display. Release
TEST pushbutton.
(4) Wait for offset values to appear on display. Offset values will be between -225
and +225.
(5) Set ammeter calibrator for 20 A output.
(6) Adjust precision oscillator frequency for 400 Hz and voltage output for 1.000 V
indication on digital multimeter.
(7) Press and release TEST pushbutton. VTM should display between 17.8 and
22.2. Record indication.
(8) Set ammeter calibrator for 100 A output.
(9) Adjust precision oscillator frequency for 400 Hz and voltage output for 1.000 V
indication on digital multimeter.
(10) Press and release TEST pushbutton. VTM should display between 93 and 107
A.
12